Optical system |
Wavelength range |
190-900nm |
Monochromator |
C-T type of eliminating aberration monochromator |
grating number |
1800lines/mm |
Spectral bandwidth |
0.1、0.2、0.4、0.7、1.0、2.0 (6 levels automatic switching) |
Wavelength accuracy |
±0.2nm |
Wavelength repeatability |
±0.1nm |
Resolution |
The spectral bandwidth can separate manganese double lines (279.5nm and 279.8nm) at 0.2nm. |
Static baseline drift |
0.002ABS/30min (Cu) |
Background correction system |
Two background subtraction methods, self-absorption and deuterium lamp, can correct 1A background |
Element lamp |
6 lamps can auto-switch |
Atomization System |
Flame analysis |
Precision |
RSD≤0.5% |
Concentration (Cu) |
0.025μg/mL/1% |
LOQ (Cu) |
0.004μg/mL |
Sprayer |
High efficiency glass sprayer |
Graphite furnace analysis |
Precision |
RSD≤3% |
Characteristic statistic (Cd) |
1.0×10-12g |
LOQ (Cd) |
2.0×10-12g |
Heating rate |
2000℃/sec |
Heating range |
3000℃ |
Safety protection |
Graphite furnace visual system, graphite furnace accidentally broken immediately stop heating and prompt alarm |
Function expansion |
Optional hydride generator for atomic absorption analysis |
Sample injection system |
Automatic sampler |
Sample quantity |
85 |
Cup material |
Polypropylene |
Number of matrix modifier |
5 |
Standard cup volume |
2mL sample,10mL reagent |
Injection volume |
1μL-50μL |
Injection accuracy |
When the volume is ≥10μL, better than 1% |
Minimum increment |
1μL |
Graphite furnace power supply |
Voltage |
220V±22V |
Power |
5000W |
Dimension |
280mm×550mm×450mm |
Main host power supply |
Voltage |
220V±22V |
Power |
200W |
Dimension |
700mm×550mm×530mm |
Reviews