Description
SPECIFICATIONS
Model | M-900 plus |
Test principle | Enzyme electrode method |
Electrode | Single replaceable electrode |
Measuring range | Glucose(Article number: C1001): 0.3 ~ 9g / L Glucose(Article number: G1001): 0.05 ~ 1g / L Lactate: 0.03 ~ 2g / L Sucrose: 0.03 ~ 2g / L Glutamate: 0.03 ~ 1g / L Glutamine: 0.03 ~ 1g / L Lysine: 0.03 ~ 1g / L Ethanol: 0.03 ~ 1g / L Methanol: 0.03 ~ 0.5g / L Glycerin: 0.03 ~ 1g / L Xylose: 0.03 ~ 1g / L The machine with pre-dilution function can expand the above test range by 10 to 30 times. NH4+:2~30mmol/L Na +:60~200mmol/L K +:0.5~15mmol/L Ca 2+:0.1~5mmol/L pH:6~8 |
Resolution | 0.01g/L; 0.01mmol/L |
Precision (CV%) | <2% |
Enzyme membrane life | Glucose: ≤30 days Lactate:≤ 30 days Glutamate: ≤30 days Glutamine: ≤15 days Lysine: ≤30 days Ethanol / methanol:≤10 days Glycerin:≤15 days Xylose:≤15 days Sucrose:≤15 days NH4+ electrode: ≥6 months Na + electrode: ≥6 months K + electrode: ≥6 months Ca 2+ electrode: ≥6 months pH electrode: ≥6 months |
Ambient temperature | 10~35°C |
Relative humidity | ≤85% (no condensation water) |
External dimension | 660×530×520mm |
Power supply | 110V or 220V, 50Hz, 60Hz, separate three-hole socket with good grounding |
Power | <100VA |
Weight | 35kg |
Screen | 10-inch color touch screen |
Reviews